Catalog Number 4004C0710 |
Device Problems
Flushing Problem (1252); Device Difficult to Setup or Prepare (1487); Failure to Advance (2524)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 10/06/2015 |
Event Type
malfunction
|
Manufacturer Narrative
|
This device is available for return.A follow up mdr will be submitted upon completion of the device investigation.
|
|
Event Description
|
During preparation for a coil embolization procedure, after removing a penumbra coil 400 coil (pc400 coil) from its packaging, the hospital staff was unable to advance the pc400 coil out of its introducer sheath and into the liquid bath (heparinized saline).The hospital staff indicated that there was no resistance on the pc400 coil pusher assembly and no reported damage to the pusher assembly.The pc400 coil was unable to be flushed for use and was not used in the procedure.The physician continued the procedure using a new pc400 coil.
|
|
Manufacturer Narrative
|
Result: the pet lock was intact on the proximal end of the penumbra coil 400 (pc400) pusher assembly and the pusher assembly was bent approximately 2.0 and 5.0 cm from the proximal end.The embolization coil was intact with the pusher assembly.Conclusion: evaluation of the returned device revealed it was able to advance and withdraw through the introducer sheath without an issue.The root cause of the complaint could not be determined.Further evaluation revealed the pusher assembly was kinked on the proximal end.This damage was likely incidental and may have occurred during packaging for return.Pc400 devices are 100% functionally evaluated during in-process inspection.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|
|
Search Alerts/Recalls
|