The customer reported there was a leak in the seal on the triple fork below the buffer spike of their clinimacs tubing set ls.Since the leak was detected prior to integrity test, no cellular material was affected and therefore no harm caused to the patient.The customer was asked to send back the tubing set for further investigation, but it had not arrived at (b)(6) at the time of this report.The customer did provide photographs, which do not have the resolution needed to identify the failure itself.Therefore, the root cause is speculative.The event happened at: (b)(6).
|