MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH UNIVERSAL IMPACTOR/POSTIONER; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/CERAMIC/POLYMER, CEMENTED OR NON-POROUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH UNIVERSAL IMPACTOR/POSTIONER; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/CERAMIC/POLYMER, CEMENTED OR NON-POROUS Back to Search Results
Catalog Number 2101-0200
Device Problems Crack (1135); Insufficient Information (3190)
Patient Problems No Consequences Or Impact To Patient (2199); No Information (3190)
Event Date 10/26/2015
Event Type  malfunction  
Manufacturer Narrative
A supplemental report will be submitted upon completion of the investigation.
 
Event Description
During a tha, the insert handle was cracked.
 
Manufacturer Narrative
An event regarding crack/fracture involving a universal impactor/positioner was reported.The event was not confirmed.Method & results: device evaluation and results: visual, dimensional, and functional inspection were not performed as the item was not returned.Medical records received and evaluation: not performed as medical records were not provided.Device history review: review of the device history records indicates devices were manufactured and accepted into final stock with no reported discrepancies.Complaint history review: there have been other events for the lot referenced.Conclusions: the event was not confirmed because the product was not returned.Previous investigations concluded that similar events for the same product and issue was concluded to be due to a design issue.Product surveillance will continue to monitor for trends.If additional information becomes available, this investigation will be reopened.
 
Event Description
During a tha, the insert handle was cracked.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNIVERSAL IMPACTOR/POSTIONER
Type of Device
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/CERAMIC/POLYMER, CEMENTED OR NON-POROUS
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer Contact
beverly lima
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key5231092
MDR Text Key31640030
Report Number0002249697-2015-03837
Device Sequence Number1
Product Code MAY
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K123604
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,other
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 10/26/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/17/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number2101-0200
Device Lot NumberB5AJS
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/18/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/19/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
-
-