MAUDE Adverse Event Report: BAXTER HEALTHCARE - CLEVELAND MINICAP; SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE - CLEVELAND MINICAP; SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE Back to Search Results
Catalog Number 5C4466P
Device Problems Crack (1135); Leak/Splash (1354)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that a minicap was cracked and leaking iodine.The product was past expiration date.The patient secured the cap to the transfer set and it resulted in an iodine leak.There was no patient injury or medical intervention associated with this event.No additional information is available.
 
Manufacturer Narrative
(b)(4).The lot was manufactured from 10/14/13 to 10/17/13.The actual device was not available; however, a companion sample was received for evaluation.Visual inspection was performed with no defects or cracks noted.Underwater pressure/leak testing was performed with no leak identified.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.The reported condition was unable to be replicated in the companion sample.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MINICAP
Type of Device
SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE
Manufacturer (Section D)
BAXTER HEALTHCARE - CLEVELAND
911 highway 61 north po box 1
cleveland MS 38732
Manufacturer (Section G)
BAXTER HEALTHCARE - CLEVELAND
911 highway 61 north po box 1
cleveland MS 38732
Manufacturer Contact
kinga almasan
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key5236837
MDR Text Key31759134
Report Number1416980-2015-42321
Device Sequence Number1
Product Code KDJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K895631
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 12/17/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/19/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date04/30/2015
Device Catalogue Number5C4466P
Device Lot NumberGD895755
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received12/09/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-