DEPUY SYNTHES POWER TOOLS BATTERY RECIPROCATOR II FOR BPL II; MOTOR, SURGICAL INSTRUMENT, AC-POWERED
|
Back to Search Results |
|
Catalog Number 530.715_LL |
Device Problems
Loss of Power (1475); Device Inoperable (1663)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 11/09/2015 |
Event Type
malfunction
|
Manufacturer Narrative
|
Additional narrative: (b)(6).As of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Event Description
|
It was reported that during sterile processing (spd process), it was discovered that the battery reciprocator device was not working.According to the reporter, when the battery pack was assembled to the reciprocator, the device had no power.A spare device was available for use.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
|
Manufacturer Narrative
|
The actual device was returned for evaluation.Reliability engineering evaluated the device and observed that the device was not working and the trigger was sticky.Therefore, the reported condition was confirmed.It was determined that an unknown liquid was found inside the unit, electronic motor and the ecu were damaged, seals and bearings were worn, and internal components were corroded.The assignable root cause was determined to seal degradation from repeated sterilization and wear on components from repeated use.If additional information should become available, a supplemental medwatch report will be sent accordingly.
|
|
Search Alerts/Recalls
|
|
|