MAUDE Adverse Event Report: DENTSPLY IMPLANTS (A DIVISION OF DENTSPLY IH AB) IMPLANT DRIVER EV 5.4 LONG; ACCESSORIES, IMPLANT, DENTAL, ENDOSSEOUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DENTSPLY IMPLANTS (A DIVISION OF DENTSPLY IH AB) IMPLANT DRIVER EV 5.4 LONG; ACCESSORIES, IMPLANT, DENTAL, ENDOSSEOUS Back to Search Results
Catalog Number 25385
Device Problems Bent (1059); Device Operates Differently Than Expected (2913)
Patient Problems Failure of Implant (1924); No Known Impact Or Consequence To Patient (2692)
Event Date 09/21/2015
Event Type  Injury  
Manufacturer Narrative
The device is available for evaluation, though results are not available as of this report.Evaluation results will be submitted as they become available.
 
Event Description
In this event a doctor reported that an ev implant driver 5.4 long was used twice and no longer will attach to a handpiece.As a result, a patient's surgery had to be aborted.
 
Manufacturer Narrative
The contra angle-lock bent at the thinnest part due to torsion, no material failure was found.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPLANT DRIVER EV 5.4 LONG
Type of Device
ACCESSORIES, IMPLANT, DENTAL, ENDOSSEOUS
Manufacturer (Section D)
DENTSPLY IMPLANTS (A DIVISION OF DENTSPLY IH AB)
aminogatan 1
molndal, vastra gotalands lan [se-14] S-431 21
SW  S-431 21
Manufacturer (Section G)
DENTSPLY IMPLANTS (A DIVISION OF DENTSPLY IH AB)
aminogatan 1
molndal, vastra gotalands lan [se-14] S-431 21
SW   S-431 21
Manufacturer Contact
helen lewis
221 w. philadelphia st.
suite 60w
york, PA 17401
7178494229
MDR Report Key5237512
MDR Text Key31671032
Report Number9612468-2015-00031
Device Sequence Number1
Product Code NDP
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation Dentist
Type of Report Initial,Followup
Report Date 10/23/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/19/2015
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number25385
Device Lot Number17403
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/23/2015
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/24/2015
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? Yes
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-