Catalog Number 2101-0200
Device Problems Break (1069); Crack (1135); Fracture (1260); Material Integrity Problem (2978)
Patient Problems No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Event Date 11/03/2015
Event Type  malfunction  
Manufacturer Narrative
A supplemental report will be submitted upon completion of the investigation.
 
Event Description
The plastic cup impactor handle cracked while putting in cup.Secondary impactor was used.
 
Manufacturer Narrative
An event regarding an alleged crack/fracture involving a universal impactor/positioner was reported.The event was confirmed.Method & results: - device evaluation and results: the returned device was in used condition.The visual inspection confirmed the handle fractured.- medical records received and evaluation: no medical records were received for review with a clinical consultant.- device history review: a review of the device history records could not be performed as no lot information was provided and the lot was illegible on the devices as there were scratches on the location of the device where the lot is located.Conclusions: the investigation concluded the handle fractured due to a design issue.
 
Event Description
The plastic cup impactor handle cracked while putting in cup.Secondary impactor was used.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNIVERSAL IMPACTOR/POSTIONER
Type of Device
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/CERAMIC/POLYMER, CEMENTED OR NON-POROUS
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer Contact
keyla navedo
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key5239317
MDR Text Key31995582
Report Number0002249697-2015-03953
Device Sequence Number1
Product Code MAY
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K123604
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 11/03/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/20/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number2101-0200
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/19/2015
Is the Reporter a Health Professional? No
Date Manufacturer Received12/17/2015
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
-
-