• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SORIN GROUP DEUTSCHLAND GMBH SORIN CENTRIFUGAL PUMP SYSTEM WITH TUBING CLAMP; CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SORIN GROUP DEUTSCHLAND GMBH SORIN CENTRIFUGAL PUMP SYSTEM WITH TUBING CLAMP; CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 60-02-15
Device Problem No Device Output (1435)
Patient Problem No Patient Involvement (2645)
Event Date 10/19/2015
Event Type  malfunction  
Manufacturer Narrative
There was no patient involvement.Device has not been returned to sorin group (b)(4).(b)(4).Sorin group (b)(4) received a report that the flow reading was showing dashes on the sorin centrifugal pump system with tubing clamp during set up.This issue occurred during priming, so there was no patient involvement.The investigation is ongoing.A follow-up report will be sent when the investigation is complete.
 
Event Description
Sorin group (b)(4) received a report that the flow reading was showing dashes on the sorin centrifugal pump system with tubing clamp during set up.This issue occurred during priming, so there was no patient involvement.
 
Manufacturer Narrative
Sorin group (b)(4) manufactures the sorin centrifugal pump system with tubing clamp.The incident occurred in (b)(6).This medwatch report is being filed on behalf of sorin group (b)(4).Sorin group (b)(4) received a report that the flow reading was showing dashes on the sorin centrifugal pump system with tubing clamp during set up.This issue occurred during priming, so there was no patient involvement.A sorin group field service representative was dispatched to the facility to investigate and was able to reproduce the reported issue.Troubleshooting identified a defective flow probe.The flow probe was replaced and no further issues were reported.A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.Sorin group (b)(4) will continue to monitor the market for trends related to this type of issue.Evaluated on site by sorin service rep.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SORIN CENTRIFUGAL PUMP SYSTEM WITH TUBING CLAMP
Type of Device
CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
SORIN GROUP DEUTSCHLAND GMBH
lindberghstr. 25
munich, 80939
GM  80939
Manufacturer (Section G)
SORIN GROUP DEUTSCHLAND
lindberghstr. 25
munich, 80939
GM   80939
Manufacturer Contact
carrie wood
14401 w. 65th way
arvada, CO 80004
3034676461
MDR Report Key5240902
MDR Text Key31788245
Report Number9611109-2015-00549
Device Sequence Number1
Product Code DWA
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K032213
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 10/19/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/20/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number60-02-15
Device Catalogue NumberN/A
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/08/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/26/2007
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-