Catalog Number 8065751511
Device Problems Scratched Material (3020); Torn Material (3024)
Patient Problem No Information (3190)
Event Type  Injury  
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event Description
An ophthalmic surgeon reported that the angled polymer tips must have had a burr on them because while using the flat surface to polish, the posterior capsules were scratched and torn.The product samples were not retained.Additional information was requested; however, none has been received to date.This is one of two reports from the same facility.
 
Manufacturer Narrative
No angled polymer tip(s) was returned for being scratched.No lot number was identified with this complaint; therefore, a lot history review could not be conducted.Because a sample was not returned and no lot information was provided for a lot record review, the root cause for customer complaint issue cannot be determined.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DISPOSABLE POLYMER IRRIGATION AND SILICONE IRRIGATION AND ASPIRATION TIPS
Type of Device
DEVICE, IRRIGATION, OCULAR SURGERY
Manufacturer (Section D)
ALCON RESEARCH, LTD. - ALCON PRECISION DEVICE
714 columbia avenue
sinking spring PA 19608
Manufacturer (Section G)
ALCON RESEARCH, LTD. - ALCON PRECISION DEVICE
714 columbia avenue
sinking spring PA 19608
Manufacturer Contact
eddie darton, md, jd
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8175686660
MDR Report Key5241132
MDR Text Key31793545
Report Number2523835-2015-00997
Device Sequence Number1
Product Code KYG
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 11/27/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/20/2015
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number8065751511
Device Lot NumberASKU
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/25/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other; Disability;
-
-