• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY SYNTHES POWER TOOLS 12.6CM TAPERED ATTACH, BM; BUR, SURGICAL, GENERAL & PLASTIC SURGERY - ATTACHMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY SYNTHES POWER TOOLS 12.6CM TAPERED ATTACH, BM; BUR, SURGICAL, GENERAL & PLASTIC SURGERY - ATTACHMENT Back to Search Results
Catalog Number B-TURQ-L
Device Problems Material Discolored (1170); Device Inoperable (1663)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/17/2015
Event Type  malfunction  
Manufacturer Narrative
(b)(6).As of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
This is report 1 of 2 for the same event.It was reported that during an unspecified neuro surgical procedure, it was observed that the motor device was not working - only blowing nitrogen and the attachment device was discolored like it got hot.It was reported that the devices were being used together.There were no delays to the surgical procedure and identical spare devices were available for use.The surgery was successfully completed.There was patient involvement reported.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Manufacturer Narrative
The actual device was returned for evaluation.Reliability engineering evaluated the device and observed that it was cosmetically unacceptable and the bearings were worn out and no longer in axial alignment creating heat.Therefore, the reported condition was confirmed.The assignable root cause was determined to be due to normal wear over time.If additional information should become available, a supplemental medwatch report will be sent accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
12.6CM TAPERED ATTACH, BM
Type of Device
BUR, SURGICAL, GENERAL & PLASTIC SURGERY - ATTACHMENT
Manufacturer (Section D)
DEPUY SYNTHES POWER TOOLS
4500 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
DEPUY SYNTHES POWER TOOLS
4500 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
rodney crain
4500 riverside drive
palm beach gardens, FL 33410
5616271080
MDR Report Key5242238
MDR Text Key32149673
Report Number1045834-2015-12574
Device Sequence Number1
Product Code GFF
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PK974025
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 11/18/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/23/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberB-TURQ-L
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/24/2015
Is the Reporter a Health Professional? No
Date Manufacturer Received12/08/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/29/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-