DEPUY SYNTHES POWER TOOLS 12.6CM TAPERED ATTACH, BM; BUR, SURGICAL, GENERAL & PLASTIC SURGERY - ATTACHMENT
|
Back to Search Results |
|
Catalog Number B-TURQ-L |
Device Problems
Material Discolored (1170); Device Inoperable (1663)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 11/17/2015 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(6).As of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Event Description
|
This is report 1 of 2 for the same event.It was reported that during an unspecified neuro surgical procedure, it was observed that the motor device was not working - only blowing nitrogen and the attachment device was discolored like it got hot.It was reported that the devices were being used together.There were no delays to the surgical procedure and identical spare devices were available for use.The surgery was successfully completed.There was patient involvement reported.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
|
Manufacturer Narrative
|
The actual device was returned for evaluation.Reliability engineering evaluated the device and observed that it was cosmetically unacceptable and the bearings were worn out and no longer in axial alignment creating heat.Therefore, the reported condition was confirmed.The assignable root cause was determined to be due to normal wear over time.If additional information should become available, a supplemental medwatch report will be sent accordingly.
|
|