Catalog Number 595000-001 |
Device Problems
Device Displays Incorrect Message (2591); Device Operates Differently Than Expected (2913)
|
Patient Problems
Vertigo (2134); No Consequences Or Impact To Patient (2199)
|
Event Date 11/15/2015 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).
|
|
Event Description
|
The reported issues involve the following syncardia temporary total artificial heart (tah-t) system components and are reported under two separate medical device reports: (1) primary freedom driver s/n (b)(4) (mfr report # 3003761017-2015-00395) and (2) freedom driver s/n (b)(4) (mfr report # 3003761017-2015-00396).The customer reported that the freedom driver s/n (b)(4) exhibited a beat rate of 122 beats per minute (bpm) though the beat rate was set at 125 bpm.Additionally, the fill volume (fv) exhibited on the driver was at 49 milliliters (ml).The customer also reported that the patient was taken to the hospital with reports of feeling a head rush.The customer also reported that the patient was subsequently switched to the backup freedom driver s/n (b)(4) at the hospital, where the driver exhibited a fault alarm.The customer also reported that the patient was subsequently switched to another freedom driver and the patient felt fine after he was switched.The freedom driver will be returned to syncardia for evaluation.The results of the investigation will be provided in a follow-up mdr.
|
|
Manufacturer Narrative
|
Corrected to patient/lay user.The freedom driver was returned to syncardia for evaluation.Visual inspection of the driver's external components revealed split housings at the bottom left side of the driver.Visual inspection of the driver's internal components revealed a broken tip on the u21 pressure sensor on the main printed circuit board assembly (pcba).Review of the electronic data revealed a fault code which is consistent with a broken u21 pressure sensor.The broken u21 pressure sensor was determined to be the root cause of the customer-reported fault alarm.Based upon the damage observed during the visual inspection including the split housings, the most likely root cause of the broken u21 pressure sensor is rough handling or an impact shock.This issue will continue to be monitored and trended as part of the customer experience process.Syncardia has completed its evaluation of this complaint and is closing this file.(b)(4).
|
|
Event Description
|
The reported issues involve the following syncardia temporary total artificial heart (tah-t) system components and are reported under two separate medical device reports: (1) primary freedom driver s/n (b)(4) (mfr report # 3003761017-2015-00395) and (2) freedom driver s/n (b)(4) (mfr report # 3003761017-2015-00396).The customer reported that the freedom driver s/n (b)(4) exhibited a beat rate of 122 beats per minute (bpm) though the beat rate was set at 125 bpm.Additionally, the fill volume (fv) exhibited on the driver was at 49 milliliters (ml).The customer also reported that the patient was taken to the hospital with reports of feeling a head rush.The customer also reported that the patient was subsequently switched to the backup freedom driver s/n (b)(4) at the hospital, where the driver exhibited a fault alarm.The customer also reported that the patient was subsequently switched to another freedom driver and the patient felt fine after he was switched.
|
|
Search Alerts/Recalls
|