User facility personnel exchanged the table with another 3085sp table and had the same issue.The table was stabilized using stackable steps, and the procedure was completed.A delay in the procedure occurred due to the exchange in tables.No injury was reported.A steris service technician arrived on site, and found that the user facility had unlocked the table in order to put the table in the desired position for the completion of the procedure.The technician tested the table with his weight and confirmed that when the table is locked it does not tip.The technician was able to get the table to tip when it was unlocked and confirmed that no malfunction occurred.The operator manual states, "warning tipping hazard: do not place patient on the table unless floor locks are engaged." as well as "do not release floor locks while patient is on the table." the table has been returned to service and no additional issues have been reported with the table.The table is under steris contract for preventative maintenance.The technician has discussed the proper use and operation of the surgical table with the user facility.
|