• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. PROXIMATE LINEAR STAPLER; STAPLE, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. PROXIMATE LINEAR STAPLER; STAPLE, IMPLANTABLE Back to Search Results
Catalog Number TX60B
Device Problem Failure to Deliver (2338)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.The lot history records were reviewed and the manufacturing criteria were met prior to the release of this lot.
 
Event Description
It was reported that during an unknown procedure, at stapling the clamp didn't deliver staples."suture was handmade".Another like device was used to complete the procedure.There were no adverse consequences for the patient.
 
Manufacturer Narrative
(b)(4).Batch # m54129.Manufacture date: 6/9/2015.Expiration date: 5/9/2020.The analysis results showed that the tx60b device arrived in good visual condition and with a reload present.The reload was received fully fired.The device was tested for functionality with a test reload and it fired and formed all the staples as intended.The device fired without any difficulties, the staples were noted to have a proper b-formation and the staple line was complete.A batch record review was performed and no anomalies were found during the manufacturing process.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROXIMATE LINEAR STAPLER
Type of Device
STAPLE, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo PR 00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC
475 calle c
guaynabo PR 00969
Manufacturer Contact
milton garrett
4545 creek road ml 120a
cincinnati, OH 45242
5133378865
MDR Report Key5246708
MDR Text Key32157272
Report Number3005075853-2015-07531
Device Sequence Number1
Product Code GDW
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K020779
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 11/01/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/24/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/09/2020
Device Catalogue NumberTX60B
Device Lot NumberM4HN1M
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/10/2015
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received12/10/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/09/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-