A review of the information in the complaint file indicates this investigation was performed by a covidien/medtronic technical center for the reported condition of; damaged power cord.Therefore, this report will be based on information provided by the technical center.The unit was triaged and the complaint was confirmed; power cord was damaged with exposed copper wires.The root cause of the power cord failure can be attributed to rough handling.Power cords periodically require replacement due to age, usage, and user damage.The power cord was replaced to correct the problem.The unit's check valve was also replaced.The unit was then fully tested; unit passed all testing.Scd 700 compression system was manufactured in 2013.A review of the device history record shows this device was released meeting all manufacturing specifications.
|