The patient was presented with a aneurysm in the right popliteal artery which was treated with the placement of a gore viabahn endoprosthesis.A 14fr flexor check-flo (cook)introducer sheath was used to advance the device over a 0.035" thruway (boston scientific) guidewire.During the deployment procedure the deployment line did not detach fully from the device.With the removal of the catheter, the already opened gore viabahn endoprosthesis was pulled back into the aneurysm.Thrombus was loosened by doing that and occluded the fibular artery and partly occluded the posterior tibial artery.A anterior tibial artery was not existing.Finally the physician was able to remove the catheter.The physician unsuccessfully tried to treat the occlusion with aspiration, thrombolysis, angioplasty and administration of prostavasin.The patient is currently being monitored.
|
Refer to field for the results of the imaging evaluation.Our engineers have evaluated the returned device.Their investigation showed following: the complete viabahn delivery system was returned to histology without the endoprosthesis.79cm of deployment line was protruding from the catheter tubing transition port upon examination of the device.Four single-fiber loops in were present in the deployment line at approximately 14cm, 15cm, 26cm, and 30cm from the frayed end of the deployment line.The deployment line had 4cm of slack at the hub.Based on the device examination performed, no manufacturing anomalies were identified.
|