Manufacturer Narrative
The field service engineer (fse) evaluated the instrument on (b)(6) 2015.The fse found that solenoid 5 was not working and was failing to actuate pinch valve pv3.The fse replaced the solenoid, which resolved the wbc issue.The repairs were verified by the fse.(b)(4).
 
Event Description
The customer reported the coulter lh 500 hematology analyzer gave no white blood cell (wbc) result for one patient sample.Erroneous patient results were not generated and there was no change or affect to patient treatment in connection with this event.
 
Manufacturer Narrative
(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COULTER LH 500 HEMATOLOGY ANALYZER
Type of Device
COUNTER, DIFFERENTIAL CELL
Manufacturer (Section D)
BECKMAN COULTER
250 south kraemer blvd.
brea CA 92821
Manufacturer (Section G)
BECKMAN COULTER
11800 sw 147th ave
miami FL 33196
Manufacturer Contact
miranda holland
11800 sw 147th ave
32-s08
miami, FL 33196
3053802031
MDR Report Key5250484
MDR Text Key32301633
Report Number1061932-2015-01763
Device Sequence Number1
Product Code GKZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K042724
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 11/03/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/25/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number178833
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/03/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/01/2007
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-