Catalog Number E3KIT
Device Problem Device Operates Differently Than Expected (2913)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
There has been a previous report received where stalling has caused file separation.Since separation of a file could necessitate medical/surgical intervention to preclude permanent damage to a body structure or permanent impairment of a body function, this malfunction would be likely to cause/contribute to a serious injury should it recur.As such, this event meets the criteria for reportability per 21 cfr part 803.The device is available for evaluation, though has not been returned as of this report.Evaluation results will be submitted as they become available.
 
Event Description
In this event it was reported that an e3 motor was running slower and feels like it is not turning the files like usual.The doctor did some trouble shooting and then reported that the motor is now stalling.No patient injury occured.
 
Manufacturer Narrative
The device has been tested for torque and speed acording to our internal standards and fullfills all the acceptance criteria.The problem was caused because attached to the device was a w&h contra-angle 16:1.It's clear that with such ratio the speed become slower while increases the torque.It is recommended to tell the dentist to read carefully the dfu and to use only sirona 6:1 contra-angles as written in the dfu.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
E3 MOTOR
Type of Device
CONTROLLER, FOOT, HANDPIECE AND CORD
Manufacturer (Section D)
DENTSPLY TULSA DENTAL SPECIALTIES
608 rolling hills drive
johnson city TN 37604
Manufacturer (Section G)
ADVANCED TECHNOLOGY RESEARCH (A.T.R.) S.R.L.
via san donato 1
pistoia, 51100
IT   51100
Manufacturer Contact
helen lewis
221 w. philadelphia st.
suite 60w
york, PA 17401
7178494229
MDR Report Key5253733
MDR Text Key32389310
Report Number9611053-2015-00053
Device Sequence Number1
Product Code EBW
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K103653
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Dentist
Type of Report Initial,Followup
Report Date 10/28/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/29/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberE3KIT
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/08/2016
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-