ANGIOMED GMBH & CO. MEDIZINTECHNIK KG FLUENCY PLUS TRACHEOBRONCHIAL STENT GRAFT Back to Search Results
Catalog Number FTM06080
Device Problem Break (1069)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
The device history records are being reviewed.The event is currently under investigation.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant was unable to provide any patient or procedural details to date.
 
Event Description
It was reported that during deployment of the endovascular stent graft, the sheath of the delivery system broke.There was no reported impact or patient injury.
 
Manufacturer Narrative
The lot history records have been reviewed with special attention to the manufacturing and inspection of this product and the product was found to have met all specifications prior to shipment.There is no indication for a manufacturing-related failure.The evaluation of the returned device confirmed the reported sheath fracture during deployment.The stent graft was found to be loaded in the system and the outer sheath was found to be torn off and elongated.The condition of the device indicates that increased friction must have affected on the delivery system during the attempt to deploy the stent graft finally resulting in the outer sheath fracture.Potential contributing factors to the reported event have been considered.Previous investigations of similar complaints have been reviewed.This type of event may be associated with a difficult vessel anatomy leading to increased friction and subsequent sheath fracture.In this case, only limited information was provided.Insufficient flushing of the device may be another contributing factor to the reported event.On the basis of the limited information available, a definite root cause for the reported event could not be determined.The ifu states that the device must be flushed with sterile saline.Also the ifu states: "the safety and effectiveness of the device when placed across a tight bend (.) has not been evaluated.Prior to stent graft deployment, ensure that the proximal (inflow) stent graft end is positioned in a straight section of the lumen to reduce the risk of higher deployment forces and possible endovascular system failure" and "do not kink the delivery catheter or use excessive force during delivery to the target lesion." furthermore, the ifu states: "if unusual resistance or high deployment force is encountered during stent graft deployment, abort the procedure, remove the delivery system and use an alternative device.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLUENCY PLUS TRACHEOBRONCHIAL STENT GRAFT
Type of Device
TRACHEOBRONCHIAL STENT GRAFT
Manufacturer (Section D)
ANGIOMED GMBH & CO. MEDIZINTECHNIK KG
wachhausstrasse 6
karlsruhe 76227
GM  76227
Manufacturer (Section G)
ANGIOMED GMBH & CO. MEDIZINTECHNIK KG -9681442
wachhausstrasse 6
karlsruhe 76227
GM   76227
Manufacturer Contact
daniela mueller
wachhausstrasse 6
karlsruhe 76227
GM   76227
0497219445
MDR Report Key5254641
MDR Text Key32415013
Report Number9681442-2015-00231
Device Sequence Number1
Product Code JCT
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K050832
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date 11/05/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/30/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/30/2017
Device Catalogue NumberFTM06080
Device Lot NumberANYF1676
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/19/2015
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/15/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/17/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-