• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) TENDRIL SDX LEAD; PERMANENT PACEMAKER ELECTRODE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) TENDRIL SDX LEAD; PERMANENT PACEMAKER ELECTRODE Back to Search Results
Model Number 1488T/46
Device Problems Break (1069); Kinked (1339)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/20/2015
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported that during a device changeout procedure, a kink was observed in the insulation of the atrial lead.The lead appeared to have been crushed.It was explanted and replaced.
 
Manufacturer Narrative
(b)(4).
 
Event Description
New information reported that noise had been observed on the lead.No patient symptoms were reported.
 
Manufacturer Narrative
Final analysis found that the lead was returned in three pieces.The outer coil was crushed, which exposed the coil as well.The damage sustained resulted from clavicular crush.The damage found likely contributed to the reported bend and noise anomalies.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TENDRIL SDX LEAD
Type of Device
PERMANENT PACEMAKER ELECTRODE
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
jaime chavez
15900 valley view court
sylmar, CA 91342
8184934022
MDR Report Key5255589
MDR Text Key32391644
Report Number2017865-2015-30558
Device Sequence Number1
Product Code NVN
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P960013
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 11/20/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/30/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/31/2007
Device Model Number1488T/46
Device Lot Number0002678497
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/01/2015
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/20/2015
Was Device Evaluated by Manufacturer? No
Date Device Manufactured08/07/2004
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age76 YR
-
-