• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOMET MICROFIXATION ELEVATOR #301

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOMET MICROFIXATION ELEVATOR #301 Back to Search Results
Model Number N/A
Device Problems Break (1069); Fracture (1260); Off-Label Use (1494)
Patient Problems No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
Current information is insufficient to permit a valid conclusion about the cause of this event.A follow up report will be sent upon completion of the device evaluation.Report two of four for the same event, see also 0001032347-2015-00420 (-1), 0001032347-2015-00475 and 0001032347-2015-00476.
 
Event Description
The facility reported they would like to return seven 301 elevators that are broken.No dates, surgery details, or patient information was able to be provided.The facility confirmed there were no adverse events.
 
Manufacturer Narrative
Upon visual inspection, the complaint has been confirmed that the tip is broken.The device history record was reviewed and no non-conformance was found for this lot.There are no indication of manufacturing defects.The ifu for this part (instructions for use) states that "the tip of the instrument is extremely thin and delicate, care should be taken to avoid applying significant pressure to the tip." the most likely underlying cause is excessive force.Supplemental report two of four from the same customer, reference 0001032347-2015-00420-2, 0001032347-2015-00475-1 and 0001032347-2015-00476-1.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELEVATOR #301
Type of Device
ELEVATOR
Manufacturer (Section D)
BIOMET MICROFIXATION
1520 tradeport drive
jacksonville FL 32218
Manufacturer (Section G)
BIOMET MICROFIXATION
1520 tradeport drive
jacksonville FL 32218
Manufacturer Contact
michelle cole
1520 tradeport drive
jacksonville, FL 32218
9047414400
MDR Report Key5256957
MDR Text Key32577795
Report Number0001032347-2015-00474
Device Sequence Number1
Product Code EMJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PEXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Dentist
Type of Report Initial,Followup
Report Date 09/10/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/01/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Dentist
Device Model NumberN/A
Device Catalogue Number09-0257
Device Lot Number082412H12
Other Device ID NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/05/2015
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/15/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/24/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Removal/Correction NumberN/A
Patient Sequence Number1
-
-