MAUDE Adverse Event Report: STRYKER ENDOSCOPY-SAN JOSE 5MM RATCHETED HANDLE 33CM; ENDOSCOPIC GRASPING/CUTTING INSTRUMENT, NON-POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ENDOSCOPY-SAN JOSE 5MM RATCHETED HANDLE 33CM; ENDOSCOPIC GRASPING/CUTTING INSTRUMENT, NON-POWERED Back to Search Results
Catalog Number 0250080234
Device Problems Device Or Device Fragments Location Unknown (2590); Misassembly by Users (3133)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/06/2015
Event Type  malfunction  
Manufacturer Narrative
Additional information will be provided once the investigation has been completed.
 
Event Description
It was reported that there were missing parts on the handle and third party insulation issues.
 
Manufacturer Narrative
The following repair diagnostic codes were identified: (broken/fractured handle).This does confirm the alleged failure mode of [missing piece on handle].Probable root cause: poor autoclave reliability, incorrect sterilization/reprocessing procedure, handling procedures, use error, shear pin failure in handle.(proximal end of device).The product was returned for investigation and the reported failure mode was confirmed.The failure mode will be monitored for future reoccurrence.
 
Event Description
It was reported that there were missing parts on the handle and third party insulation issues.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
5MM RATCHETED HANDLE 33CM
Type of Device
ENDOSCOPIC GRASPING/CUTTING INSTRUMENT, NON-POWERED
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer (Section G)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer Contact
kimberly lynch
5900 optical court
san jose, CA 95138
4087542000
MDR Report Key5257418
MDR Text Key32490653
Report Number0002936485-2015-01084
Device Sequence Number1
Product Code OCZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K973259
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 11/06/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/01/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number0250080234
Device Lot Number11062015
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/24/2015
Is the Reporter a Health Professional? No
Date Manufacturer Received11/06/2015
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-