• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CODMAN & SHURTLEFF, INC. / MEDOS S.A. HAKIM INLINE PROGRAMMABLE VALVE SG; SHUNT, CENTRAL NERVOUS SYSTEM & COMPS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CODMAN & SHURTLEFF, INC. / MEDOS S.A. HAKIM INLINE PROGRAMMABLE VALVE SG; SHUNT, CENTRAL NERVOUS SYSTEM & COMPS Back to Search Results
Catalog Number 82-3162
Device Problem Device Issue (2379)
Patient Problem Headache (1880)
Event Date 11/09/2015
Event Type  Injury  
Manufacturer Narrative
(b)(4).Upon completion of the investigation a follow up report will be filed.
 
Event Description
Valve set at 200 mmh20.Patient had severe headaches in positional changes.Flow study ok.Maybe sg issue? explanted and implanted new valve and patient is doing better.
 
Manufacturer Narrative
Upon completion of the investigation it was noted that the position of the cam when valve was received was 200mmh2o.The valve was visually inspected, needle holes in the needle chamber were noted.The valve was irrigated with purified water.No occlusion was noted.The siphon guard was irrigated with purified water, no occlusions were noted.The valve was dried.The valve was leak tested, only leaked from the needle holes in the needle chamber.The valve was tested for programming.The valve passed the test.The siphon guard was removed.The valve was then pressure tested, the valve passed the test.Review of the history device records confirmed the valve product code 82-3162 with lot crbdb4, conformed to the specifications when released to stock in 17th march 2014.No root cause could be determined as the valve functions, no bend could be found with the valve.Based on the results of this investigation no further action is required.Trends will be monitored for this and similar complaints.At the present time this complaint is closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HAKIM INLINE PROGRAMMABLE VALVE SG
Type of Device
SHUNT, CENTRAL NERVOUS SYSTEM & COMPS
Manufacturer (Section D)
CODMAN & SHURTLEFF, INC. / MEDOS S.A.
rue girardet 29
ch-2400 le locle
SZ 
Manufacturer Contact
james kenney
325 paramount drive
raynham, MA 02767
5088282726
MDR Report Key5258904
MDR Text Key32450792
Report Number1226348-2015-10717
Device Sequence Number1
Product Code JXG
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PK992173
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/01/2015
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date01/31/2019
Device Catalogue Number82-3162
Device Lot NumberCRBDB4
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/04/2015
Is the Reporter a Health Professional? No
Date Manufacturer Received12/08/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/17/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-