SORIN GROUP DEUTSCHLAND GMBH SORIN CENTRIFUGAL PUMP SYSTEM WITH TUBING CLAMP; CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 60-02-15
Device Problem No Display/Image (1183)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/07/2015
Event Type  malfunction  
Manufacturer Narrative
Sorin group (b)(4) manufactures the sorin centrifugal pump system with tubing clamp.The incident occurred in (b)(6).This medwatch report is being filed on behalf of sorin group (b)(4).Sorin group (b)(4) received a report that the flow was not correctly displayed on the sorin centrifugal pump system with tubing clamp during a procedure.There was no report of patient injury.The investigation is ongoing.A follow-up report will be sent when the investigation is complete.
 
Event Description
Sorin group (b)(4) received a report that the flow was not correctly displayed on the sorin centrifugal pump system with tubing clamp during a procedure.There was no report of patient injury.
 
Manufacturer Narrative
Sorin group (b)(4) manufactures the sorin centrifugal pump system with tubing clamp.The incident occurred in (b)(6).This medwatch report is being filed on behalf of sorin group (b)(4).Sorin group (b)(4) received a report that the flow was not correctly displayed on the sorin centrifugal pump system with tubing clamp during a procedure.There was no report of patient injury.A sorin group field service representative was dispatched to the facility to investigate and was able to confirm the reported issue.Troubleshooting identified a defective flow board.The flow board was replaced and the unit was tested.No further issues were reported.A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.Sorin group (b)(4) will continue to monitor the market for trends related to this type of issue.Evaluated on site by sorin service rep.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SORIN CENTRIFUGAL PUMP SYSTEM WITH TUBING CLAMP
Type of Device
CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
SORIN GROUP DEUTSCHLAND GMBH
lindberghstr. 25
munich, 80939
GM  80939
Manufacturer (Section G)
SORIN GROUP DEUTSCHLAND
lindberghstr. 25
munich, 80939
GM   80939
Manufacturer Contact
carrie wood
14401 w. 65th way
arvada, CO 80004
3034676461
MDR Report Key5259214
MDR Text Key32692154
Report Number9611109-2015-00579
Device Sequence Number1
Product Code DWA
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K032213
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 11/07/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/01/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number60-02-15
Device Catalogue NumberN/A
Device Lot NumberN/A
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/08/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/25/2008
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-