Catalog #905103, lot # 07101507.Method: visual inspection; functional inspection; device history review; complaint history review; risk assessment; results: manufacturing records were reviewed for the corresponding lot and no relevant issues were identified.Upon visual inspection, tubing set was confirmed to be slightly bent, this could have been caused by applying extensive force by the user.No damage to the tubing set was noticed prior the surgery.Conclusion: the root cause of the tubing set leakage is due to the cut on the distal end of the tube and is likely user related.It is unclear as to when this cut occurred.
|