• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER INSTRUMENTS-KALAMAZOO UNIVERSAL HANDSWITCH; UNIT, OPERATIVE DENTAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER INSTRUMENTS-KALAMAZOO UNIVERSAL HANDSWITCH; UNIT, OPERATIVE DENTAL Back to Search Results
Catalog Number 5100009000
Device Problem Positioning Problem (3009)
Patient Problem No Patient Involvement (2645)
Event Date 11/05/2015
Event Type  malfunction  
Event Description
The handswitch was returned to stryker instruments for service.During functional testing by a service technician, it was found that the run/safe switch does not lock into either position, presenting a potential for the device to inadvertently be activated.No patient involvement, no delay, no medical intervention, and no adverse consequences were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNIVERSAL HANDSWITCH
Type of Device
UNIT, OPERATIVE DENTAL
Manufacturer (Section D)
STRYKER INSTRUMENTS-KALAMAZOO
4100 east milham avenue
kalamazoo MI 49001
Manufacturer (Section G)
STRYKER INSTRUMENTS-KALAMAZOO
4100 east milham avenue
kalamazoo MI 49001
Manufacturer Contact
casey metzger
4100 east milham avenue
kalamazoo, MI 49001
2693237700
MDR Report Key5260769
MDR Text Key32493759
Report Number0001811755-2015-04379
Device Sequence Number1
Product Code EIA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K032117
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial
Report Date 11/05/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/02/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number5100009000
Device Lot Number12129
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/04/2015
Date Manufacturer Received11/05/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/08/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-