The investigation determined that lower than expected vitros glu results were obtained from vitros performance verifier controls processed using vitros glucose (glu) slides on a vitros 250 chemistry system.The investigation concluded that the likely cause of the lower than expected qc results was a suboptimal glu calibration.However, the cause of the suboptimal glu calibration was unknown.An unknown issue with the vitros calibrator fluids or the glu cartridge in use during the calibration events could not be ruled out as potential contributing factors.There was no evidence to suggest the vitros glu reagent of vitros 250 system malfunctioned.
|