ST. JUDE MEDICAL, INC.(CRM-SYLMAR) VERITY ADX XL; IMPLANTABLE PACEMAKER PULSE GENERATOR Back to Search Results
Model Number 5056
Device Problems Difficult to Interrogate (1331); Pacemaker Found in Back-Up Mode (1440)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/20/2015
Event Type  Injury  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported that the pulse generator could not be interrogated.Communication was established and the device was found to be operating in backup vvi mode.Battery depletion was found to have occurred.It was noted that the patient had never been seen for follow-up.The device was explanted and replaced on an unknown date.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VERITY ADX XL
Type of Device
IMPLANTABLE PACEMAKER PULSE GENERATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
jaime chavez
15900 valley view court
sylmar, CA 91342
8184934022
MDR Report Key5262752
MDR Text Key32577222
Report Number2017865-2015-30132
Device Sequence Number1
Product Code LWP
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
P880086
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 10/20/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/02/2015
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/30/2014
Device Model Number5056
Device Lot Number3902426
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/10/2015
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/20/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/21/2012
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-