HILL-ROM BATESVILLE AFFINITY 4 BED FRAME; BIRTHING BED Back to Search Results
Model Number P3700D000063
Device Problem Self-Activation or Keying (1557)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/05/2015
Event Type  malfunction  
Manufacturer Narrative
The hill-rom technician found the left hand side rail needed to have the contacts cleaned.Per the hill-rom service manual the affinity® three birthing bed and affinity® four birthing beds require an effective maintenance program.We recommend that you perform semiannual preventive maintenance.Check the switches in the side rails to ensure they are functioning correctly.Also check for intermittent operation.A search of the hill-rom maintenance records did not show hill-rom performed any preventative maintenance on this bed.It is unknown if the facility performs preventative maintenance on their beds.The technician cleaned the left hand side rail contacts and the issue was resolved.Based on this information, no further action is required.
 
Event Description
Hill-rom received a report from the account stating the head section would self-run when the bed is plugged in.The bed was located in the labor ward at the account.There was no patient/user injury reported.This report was filed in our complaint handling system as complaint #(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AFFINITY 4 BED FRAME
Type of Device
BIRTHING BED
Manufacturer (Section D)
HILL-ROM BATESVILLE
1069 state route 46 east
batesville IN 47006
Manufacturer Contact
tony werner
1069 state route 46 east
batesville, IN 47006
8129312359
MDR Report Key5265339
MDR Text Key32809724
Report Number0001824206-2015-01059
Device Sequence Number1
Product Code KNC
Combination Product (y/n)N
PMA/PMN Number
K915779
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other
Type of Report Initial
Report Date 11/05/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/03/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other Caregivers
Device Model NumberP3700D000063
Was Device Available for Evaluation? Yes
Date Manufacturer Received11/05/2015
Was Device Evaluated by Manufacturer? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-