Model Number ICONPBN-HT
Device Problem Cut In Material (2454)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Method: the complaint icon cpap was received at fisher and paykel healthcare in (b)(4) for evaluation.The icon cpap was inspected externally.Results: visual inspection revealed no signs of impact damage to the returned cpap.The power cord was observed to be cut.The device's hour meter read 1,638 hours, indicating it had been used for approximately 200 days.A lot check revealed no other complaints of damaged power cords for lot number 140806.Conclusion: from the investigation conducted, the nature of the damage observed on the power cord suggests that it was cut.It is unknown how the cord became cut, however based on the number of hours used before a complaint was noted, the cut most likely occurred when in use by the customer.During initial assembly of the icon cpap, all power cords are visually inspected for damage.Once the assembly process is completed, all icon cpap devices are visually inspected again before release for distribution.This suggests the damage occured after it had been distributed.The icon cpap is designed to the electrical safety standards ,ul60601-1 and as/nzs 3200.1.The materials used in the thermoplastic components of the mains connector and the cases are flame retardant according to ul 60601-1 and as/nzs 3200.1.Our user instructions that accompany the icon state the following: only operate if the device, power cord and plug are dry and in good working order.Do not operate the device, water chamber or breathing tube if it is dropped, damaged or not working as intended.
 
Event Description
A healthcare facility reported that the power cord of an icon cpap humidifier was coming out the back of the machine and will not power on.No patient consequence was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ICON SERIES CPAP
Type of Device
BZD
Manufacturer (Section D)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ  2013
Manufacturer (Section G)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ   2013
Manufacturer Contact
raymond yan
15365 barranca parkway
irvine, CA 92618-2216
9194534000
MDR Report Key5266247
MDR Text Key32662314
Report Number9611451-2015-00515
Device Sequence Number1
Product Code BZD
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K094040
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Respiratory Therapist
Type of Report Initial
Report Date 11/06/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/04/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Respiratory Therapist
Device Model NumberICONPBN-HT
Device Catalogue NumberICONPBN-HT
Device Lot Number140806
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/24/2015
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/06/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/06/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-