Actual device was returned for investigation on 09/22/2015.Device not being used on patient when issues occured.Rwmic received device, with tip off, and upon visual inspection found a small amount of solder.Amount of solder was insufficient in order to maintain an intact device during normal use.Manufacturing records revealed one individual performed the soldering step for this device lot.Manufacturing instructions to be reviewed with individual to assure manufacturing steps are clearly understood.Remaining inventory to be visually inspected as well as stress tested to verify sufficient amount of solder used on any remaining devices.One similar issue has occured, mdr1418479-2015-00041.Rwmic considers this report closed.
|