DEPUY SYNTHES POWER TOOLS UNIVERSAL BATTERY CHARGER II; BATTERY, REPLACEMENT, RECHARGEABLE
|
Back to Search Results |
|
Catalog Number 05.001.204 |
Device Problems
Failure to Charge (1085); Defective Component (2292); Failure of Device to Self-Test (2937)
|
Patient Problem
No Patient Involvement (2645)
|
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(6).This device was returned for service; however, did not meet manufacturing specifications during pre-repair assessment.Reliability engineering evaluated the device and observed that the blue led light was flashing, failed self test, charger was defective and did not charge.Therefore, the reported condition was confirmed.However, an assignable root cause was not determined.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Event Description
|
It was reported by (b)(6) that during service and evaluation, it was observed that the blue led light was flashing, failed self test, charger was defective and did not charge on the universal battery charger device.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of this event was unknown.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
|