• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CAREFUSION ALARIS® PUMP MODULE ADMINISTRATION SET; SET, ADMINISTRATION, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CAREFUSION ALARIS® PUMP MODULE ADMINISTRATION SET; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Model Number 2420-0500
Device Problems Crack (1135); Material Disintegration (1177)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
The affected product has been received and the evaluation is pending.A follow up report will be submitted once the evaluation is completed.
 
Event Description
The customer reported that the tubing failed due to cracks and disintegration at the connection site to the patient.There was no patient harm reported.
 
Manufacturer Narrative
The customer¿s report of a damaged tubing component was confirmed.Visual inspection of the primary set¿s distal luer lock found that the spin collar was cracked.Functional testing was not performed due to the obvious damage.Analysis and testing of component samples from similar reports has determined the damage was likely due to a brittle failure.The root cause of the damaged tubing component was identified as a defective luer collar component.A manufacturing corrective action and field action have been implemented.
 
Manufacturer Narrative
Correction: initial reporter address.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALARIS® PUMP MODULE ADMINISTRATION SET
Type of Device
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
CAREFUSION
10020 pacific mesa blvd
san diego CA 92121 4386
Manufacturer (Section G)
CAREFUSION
10020 pacific mesa blvd
san diego CA 92121 4386
Manufacturer Contact
sylvia ventura
10020 pacific mesa blvd
san diego, CA 92121-4386
8586172000
MDR Report Key5272048
MDR Text Key32861228
Report Number9616066-2015-01776
Device Sequence Number1
Product Code FPA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K944320
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,other
Reporter Occupation Other
Remedial Action Recall
Type of Report Initial,Followup,Followup
Report Date 11/17/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/07/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number2420-0500
Device Catalogue Number2420-0500
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/02/2015
Is the Reporter a Health Professional? No
Date Manufacturer Received04/22/2016
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-