Model Number 5392 |
Device Problem
Connection Problem (2900)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 09/10/2015 |
Event Type
malfunction
|
Manufacturer Narrative
|
The information submitted reflects all relevant data received.If additional relevant information is received, a supplemental report will be submitted.Product event summary: analysis confirmed that the output connector assembly was broken.It was also noted that the main printed circuit board was out of electrical specification.(b)(4).
|
|
Event Description
|
It was reported that the atrial and ventricular connectors of the external pulse generator were broken.The epg was returned for repair.There was no patient involvement.
|
|
Manufacturer Narrative
|
Further analysis was performed on the main board.Visual inspection revealed no anomalies.Bench analysis found that all tests passed.The error log was reviewed, there were no failures in the log.Conclusion: the main board failure could not be verified.
|
|