• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAQUET CARDIOPULMONARY AG MAQUET HCU40 DEVICE; CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET CARDIOPULMONARY AG MAQUET HCU40 DEVICE; CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 70104.4054
Device Problem Break (1069)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The unit is under investigation.Currently no results available.Please note: the valve is responsible for heating/cooling the water flow.A supplemental medwatch will be submitted as soon as additional information becomes available.
 
Event Description
"on the hcu40 serial number (b)(4) was a broken pin inside the cplg mixing valve." no patient was involved.The malfunction was noticed during maintenance.(b)(4).
 
Manufacturer Narrative
(b)(4).A maquet field service technician investigated the unit.The technician found a broken pin inside the mixing valve.The mixing valve was replaced and the unit was tested for functionality and to factory specifications.All tests were performed successfully.A supplemental medwatch will be submitted if additional information becomes available.
 
Event Description
(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAQUET HCU40 DEVICE
Type of Device
CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
MAQUET CARDIOPULMONARY AG
rastatt
GM 
Manufacturer (Section G)
BERND RAKOW
maquet cardiopulmonary ag
kehler strasse 31
76437 rastatt 76437
GM   76437
Manufacturer Contact
maquet cardiopulmonary ag
kehler strasse 31
76437 rastatt 76437
4972229321
MDR Report Key5274848
MDR Text Key33431972
Report Number8010762-2015-01220
Device Sequence Number1
Product Code DWC
Combination Product (y/n)N
Reporter Country CodeNL
PMA/PMN Number
K130300
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 03/21/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/08/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model Number70104.4054
Device Catalogue Number70104.4054
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/10/2015
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/21/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/01/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-