Catalog Number PVPS
Device Problem Torn Material (3024)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/19/2015
Event Type  malfunction  
Manufacturer Narrative
(b)(4).To date, the device has not been returned.If the product is returned for evaluation, any further info derived from the evaluation will be submitted in a supplemental 3500a form.
 
Event Description
It was reported that the patient underwent an umbilical hernia procedure on (b)(6) 2015 and mesh was implanted.During the placement, it was noticed that the mesh was torn down.It was also reported that the procedure was completed and there were no adverse patient consequences.
 
Manufacturer Narrative
In addition, a review of the batch manufacturing records was conducted and the batch met all finished goods release criteria.
 
Manufacturer Narrative
Conclusion: the product upon which this medwatch is based has been received, however, the product evaluation is not yet complete.Any further information derived from the evaluation will be submitted in a supplemental 3500a form.
 
Manufacturer Narrative
The actual sample was returned for evaluation.Visual examination revealed the partial separation of the wings at the welding with the load ring, which may occur during handling and does not have any effect on the product properties and functionality.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROCEED VENTRAL PATCH
Type of Device
MESH, SURGICAL, POLYMERIC
Manufacturer (Section D)
ETHICON INC.
p.o. box 151, route 22 west
somerville NJ 08876 0151
Manufacturer (Section G)
ETHICON INC.-GMBH
robert-koch strasse 1
norderstedt D-228 51
GM   D-22851
Manufacturer Contact
guillermo villa
route 22 west po box 151
somerville, NJ 08876
9082180707
MDR Report Key5275131
MDR Text Key32969158
Report Number2210968-2015-19864
Device Sequence Number1
Product Code FTL
Combination Product (y/n)N
PMA/PMN Number
K061533
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial,Followup,Followup,Followup
Report Date 11/20/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/08/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/30/2017
Device Catalogue NumberPVPS
Device Lot NumberJH8DQWA0
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/27/2015
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received01/04/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/01/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-