• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PLEXUS MANUFACTURING SDN. BHD; PULSE-GENERATOR, PACEMAKER, EXTERNAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PLEXUS MANUFACTURING SDN. BHD; PULSE-GENERATOR, PACEMAKER, EXTERNAL Back to Search Results
Model Number 5392
Device Problems Nonstandard Device (1420); Device Displays Incorrect Message (2591)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/08/2015
Event Type  malfunction  
Manufacturer Narrative
This event occurred outside the us where the same model is distributed.All information provided is included in this report.Patient information is not generally available due to confidentiality concerns.Product event summary: analysis confirmed the reported event, the main printed circuit board (pcb) was out of electrical specification.It was also noted that the upper case was broken.(b)(4).
 
Event Description
It was reported that an error code was displayed.The device was returned for repair.No patient complications have been reported as a result of this event.
 
Manufacturer Narrative
Further analysis was performed on the main printed circuit board assembly (pcba).Visual inspection revealed no anomalies.Bench analysis found that the device powered up normally.The device was run on the automated test console, all tests passed.The unit was tested hot and cold, could not reproduce the error.The log was reviewed and the reported errors were verified in the log file.Conclusion: the errors were verified in the log file, however could not reproduce the error in the lab.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Type of Device
PULSE-GENERATOR, PACEMAKER, EXTERNAL
Manufacturer (Section D)
PLEXUS MANUFACTURING SDN. BHD
bayan lepas free industrial zo
bayan lepas 11900
MY  11900
Manufacturer (Section G)
MEDTRONIC CARDIAC RHYTHM HEART FAILURE
8200 coral sea st ne
mounds view MN 55112
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key5283552
MDR Text Key33508815
Report Number3004593495-2015-00436
Device Sequence Number1
Product Code DTE
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
K150246
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation Biomedical Engineer
Remedial Action Recall
Type of Report Initial,Followup
Report Date 10/08/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/09/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number5392
Device Catalogue Number5392
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/15/2015
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/08/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/17/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberZ-1356-2015
Patient Sequence Number1
-
-