DEPUY SYNTHES POWER TOOLS UNIV-BATTERY CHARGER F/SYNTHES BATT 4 CH; BATTERY, REPLACEMENT, RECHARGEABLE
|
Back to Search Results |
|
Catalog Number 530.601 |
Device Problems
Corroded (1131); Device Inoperable (1663)
|
Patient Problem
No Patient Involvement (2645)
|
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(6).The device was returned for service; however, did not meet manufacturing specifications during pre-repair assessment.Reliability engineering evaluated the device and found that the universal battery charger device was not functioning.Therefore, the reported condition was confirmed.It was further determined that the led on the left loading bay was not lit after switching the device on and the contact was corroded.The assignable root cause was determined to be due to strain/wear from normal use and servicing over time.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
|
Event Description
|
It was reported by (b)(6) that during service and evaluation, it was observed that the universal battery charger device was not functioning.It was reported that the led on the left loading bay was not lit after switching the device on and the contact was corroded.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of this event was unknown.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
|
Search Alerts/Recalls
|
|
|