DEPUY SYNTHES POWER TOOLS UNIV-BATTERY CHARGER F/SYNTHES BATT 4 CH; BATTERY, REPLACEMENT, RECHARGEABLE Back to Search Results
Catalog Number 530.601
Device Problems Corroded (1131); Device Inoperable (1663)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
(b)(6).The device was returned for service; however, did not meet manufacturing specifications during pre-repair assessment.Reliability engineering evaluated the device and found that the universal battery charger device was not functioning.Therefore, the reported condition was confirmed.It was further determined that the led on the left loading bay was not lit after switching the device on and the contact was corroded.The assignable root cause was determined to be due to strain/wear from normal use and servicing over time.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Event Description
It was reported by (b)(6) that during service and evaluation, it was observed that the universal battery charger device was not functioning.It was reported that the led on the left loading bay was not lit after switching the device on and the contact was corroded.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of this event was unknown.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNIV-BATTERY CHARGER F/SYNTHES BATT 4 CH
Type of Device
BATTERY, REPLACEMENT, RECHARGEABLE
Manufacturer (Section D)
DEPUY SYNTHES POWER TOOLS
hauptstrasse 24
waldenburg CH443 7
SZ  CH4437
Manufacturer (Section G)
DEPUY SYNTHES POWER TOOLS
hauptstrasse 24
waldenburg CH443 7
SZ   CH4437
Manufacturer Contact
linda plews
1302 wrights lane east
west chester, PA 19380
6107195000
MDR Report Key5286949
MDR Text Key33826642
Report Number8030965-2015-12147
Device Sequence Number1
Product Code MOQ
Combination Product (y/n)N
Reporter Country CodeSZ
PMA/PMN Number
PEXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Service and Testing Personnel
Type of Report Initial
Report Date 11/16/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/10/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number530.601
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/18/2015
Is the Reporter a Health Professional? No
Date Manufacturer Received11/23/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/23/2008
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-