(b)(4).The complaint airvo humidifier was returned to our fisher & paykel healthcare (fph) regional office in (b)(4) where it was inspected by a trained fph technician.The device was performance tested at the customer settings and with the same model of flow meter.The subject airvo passed all functional tests and was able to achieve 100% fio2.The hospital had stated that the airvo had displayed and sounded an alarm but were not able to tell us which alarm.Based on the description of events it is possible that the flow had become restricted due to a blockage in the breathing system.The most likely cause for this is kinking of the prongs.In this event the airvo will alarm with a "blockage" audio and visual alarm.The user should then check for any blokages in order to restore normal operation.No fault was found with the subject airvo.We have concluded that the reported incident was not caused by a malfunction of the airvo itself but possibly by a blockage in the flow delivery system, thereby restricting flow to the patient and leading to desaturation.The airvo 2 user manual provides the folllowing warnings: - do not block the flow of air through the unit and breathing tube - check the heated breathing tube or patient interface for blockage - the oxygen concentration delivered to the patient can be affected by changes to the flow setting, oxygen setting, patient interface or if the airpath is obstructed.The manual also states that the "airvo is for the treatment of spontaneously breathing patients who would benefit from receiving high-flow, warmed and humidified respiratory gases." and that "the unit is not intended for life support.".
|