• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY SYNTHES POWER TOOLS 5.0 CM SHORT ATTACHMENT; MOTOR, DRILL, ELECTRIC - ATTACHMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY SYNTHES POWER TOOLS 5.0 CM SHORT ATTACHMENT; MOTOR, DRILL, ELECTRIC - ATTACHMENT Back to Search Results
Catalog Number SHORT
Device Problems Break (1069); Component Missing (2306); Material Distortion (2977); Mechanical Jam (2983)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
(b)(6).This device was returned for service; however, did not meet manufacturing specifications during pre-repair assessment.Reliability engineering evaluated the device and the reported condition was confirmed.The assignable root cause was determined to be due to wear from normal use and servicing over time.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported from (b)(6) that during service and evaluation, it was observed that the bearings on the attachment device were damaged.It was further determined that the bearings were defective, tool was jammed, pins of the extension shaft were sheared and distorted, and the triangle symbol was missing.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of this event was unknown.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Manufacturer Narrative
The serial number of the device was inadvertenly entered into the lot number field instead of the serial number field in the initial report.The serial number of the device was removed from the lot number field and entered into the serial number field to reflect the correct information.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
5.0 CM SHORT ATTACHMENT
Type of Device
MOTOR, DRILL, ELECTRIC - ATTACHMENT
Manufacturer (Section D)
DEPUY SYNTHES POWER TOOLS
4500 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
DEPUY SYNTHES POWER TOOLS
4500 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
rodney crain
4500 riverside drive
palm beach gardens, FL 33410
5616271080
MDR Report Key5288754
MDR Text Key33950620
Report Number1045834-2015-12744
Device Sequence Number1
Product Code HBC
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PK011444
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Service and Testing Personnel
Type of Report Initial,Followup
Report Date 11/03/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/11/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberSHORT
Device Lot Number31182817
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/19/2015
Is the Reporter a Health Professional? No
Date Manufacturer Received12/30/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/17/2006
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-