MAKO SURGICAL CORP. MCK TIBIAL ONLAY INSERT-SZ 4-8MM; PROSTHESIS, KNEE PATELLOFEMOROTIBIAL, PARTIAL, SEMI-CONSTRAINED, CEMENTED, POLYM
|
Back to Search Results |
|
Catalog Number 180704-1 |
Device Problems
Break (1069); Fracture (1260); Material Integrity Problem (2978); Appropriate Term/Code Not Available (3191)
|
Patient Problems
Bone Fracture(s) (1870); Injury (2348); No Code Available (3191)
|
Event Date 11/16/2015 |
Event Type
Injury
|
Manufacturer Narrative
|
When completed, the investigation results will be submitted in a supplemental report.
|
|
Event Description
|
Revision poly insert swap.
|
|
Manufacturer Narrative
|
An event regarding crack/fracture involving a mako insert was reported.The event was confirmed.Method & results: device evaluation: visual inspection of the returned device confirmed that the device is cracked.Machine markings are still present.Medical records received and evaluation: no medical records were received for review with a clinical consultant.Device history review indicated the devices accepted into final stock from the reported lot were free from discrepancies.Complaint history review: there have been no other events for the lot referenced.Conclusion: the investigation confirmed that the device was damaged.There is no indication at this time that the design, materials, or manufacturing of the subject device contributed to the event.If additional information becomes available, this investigation will be reopened.
|
|
Event Description
|
Revision poly insert swap.Update: the received device was fractured.
|
|