SYNTHES ELMIRA TI OBA PLATE ANCHOR DOMED DESIGN 4 HOLES; IMPLANT, ENDOSSEOUS, ORTHODONTIC Back to Search Results
Catalog Number 04.500.013
Device Problem Device Operates Differently Than Expected (2913)
Patient Problem Failure of Implant (1924)
Event Type  Injury  
Manufacturer Narrative
Device was used for treatment, not diagnosis.Unknown when device malfunctioned.(b)(4).Device is expected to be returned for manufacturer review/investigation, but has yet to be received.Subject device has not been received.Without a lot number, the device history record review and the investigation could not be completed; no conclusion could be drawn, as no product was received.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that a surgeon performed a bone-anchored maxillary protraction (bamp) procedure and implanted orthodontic bone anchor (oba) implants.The patient came in for follow-up approximately one month ago.The surgeon noticed one of the oba implants was loose.The surgeon removed loose hardware on (b)(6) 2015 and put a new plate in.The patient is doing well.No surgical delay reported.This report is 4 of 4 for (b)(4).
 
Manufacturer Narrative
Lot number provided by reporter.Date returned to manufacturer.Subject device has been received and is currently in the evaluation process.Investigation is ongoing; no conclusion could be drawn as product is entering the complaint system.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Product manufacture date: november 5, 2014.A review of the plate device history record (dhr) revealed no complaint related anomalies.The (dhr) shows this lot of titanium oba plate anchor domed - 4 hole was processed through the normal manufacturing and inspection operations with no non-conformances or rework noted.This order met all dimensional and visual criteria at the time of release with no issues documented during its manufacture that would contribute to this complaint condition a product investigation was completed: the following device(s) was received: anchor t-plate, 4 hole, domed head, orthodontic bone anchor system (part 04.500.013, lot 7798054) and 1.55mm midface self-drilling screw (part 04.500.026.01, lot unknown, quantity: 3).The returned devices are in good condition.Both returned plates have been contoured and cut, with minor cosmetic damage which indicates that the plates were once implanted.The returned screws are in good condition with no functional damage and only minor marring around the hex drive.Due to the nature of the complaint, the complaint condition was unable to be replicated or verified.A visual inspection, complaint history review, drawing review, and risk assessment review were performed as part of this investigation.During the investigation no product design issues or discrepancies were observed that may have contributed to the complaint condition.A visual inspection, complaint history review, drawing review, and risk assessment review were performed as part of this investigation.No product design issues or discrepancies were observed.This complaint is unconfirmed.The relevant drawings for the device(s) were reviewed.No drawing issues or discrepancies were noted.The design was determined to be suitable for the intended use when employed and maintained as recommended.Proper use and maintenance for the device(s) are addressed in the technique guide.The root cause of the complaint condition is unknown.No manufacturing or design issues were noted during the investigation.The design determined to be adequate for its intended use when used and maintained as recommended device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TI OBA PLATE ANCHOR DOMED DESIGN 4 HOLES
Type of Device
IMPLANT, ENDOSSEOUS, ORTHODONTIC
Manufacturer (Section D)
SYNTHES ELMIRA
35 airport road
horseheads NY 14845
Manufacturer (Section G)
SYNTHES ELMIRA
35 airport road
horseheads NY 14845
Manufacturer Contact
linda plews
1302 wrights lane east
west chester, PA 19380
6107195000
MDR Report Key5299457
MDR Text Key33582622
Report Number2520274-2015-17886
Device Sequence Number1
Product Code OAT
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PK093299
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 11/24/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/15/2015
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number04.500.013
Device Lot Number7798054
Other Device ID Number(01)10887587016528(10)7798054
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/15/2015
Is the Reporter a Health Professional? No
Date Manufacturer Received01/11/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/05/2014
Is the Device Single Use? Yes
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age13 YR
-
-