• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEXCOM, INC. DEXCOM G5 MOBILE CONTINUOUS GLUCOSE MONITORING SYSTEM; MDS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEXCOM, INC. DEXCOM G5 MOBILE CONTINUOUS GLUCOSE MONITORING SYSTEM; MDS Back to Search Results
Model Number 9438-06
Device Problem Application Program Problem (2880)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/21/2015
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
Patient contacted dexcom on (b)(6) 2015, to report that on (b)(6) 2015, the g5 transmitter could not be found on the g5 application.No additional event or patient information is available.The device has been received.Investigation is not yet complete.A follow-up will be submitted upon completion of device analysis.
 
Manufacturer Narrative
(b)(4).
 
Event Description
The complaint device was returned for evaluation.The device was visually inspected and no defect was found.Sensor voltage was measured and failed.The transmitter was tested on a transmitter tester and was unable to connect resulting in failure.The customer complaint was confirmed.The root cause was determined to be a defective transmitter.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DEXCOM G5 MOBILE CONTINUOUS GLUCOSE MONITORING SYSTEM
Type of Device
MDS
Manufacturer (Section D)
DEXCOM, INC.
6340 sequence drive
san diego CA 92121
Manufacturer (Section G)
DEXCOM, INC.
6340 sequence drive
san diego CA 92121
Manufacturer Contact
kipp durbin
6340 sequence drive
san diego, CA 92121
8582000200
MDR Report Key5299804
MDR Text Key34122195
Report Number3004753838-2015-82491
Device Sequence Number1
Product Code MDS
UDI-Device Identifier00386270000224
UDI-Public00386270000224
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P120005
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 11/21/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/15/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number9438-06
Device Catalogue NumberSTT-GF-001
Device Lot Number5206822
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/07/2015
Is the Reporter a Health Professional? No
Date Manufacturer Received12/30/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/13/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age25 YR
-
-