(b)(4).Device evaluated by manufacturer - the device has a kink at 15mm from the tip while in the neutral position between ring 1 and 2.In addition the ring #1, #2 & #3 have broken adhesive and fluids under them.The steering knob and the tension control knob functioned properly on both lock and unlock positions.Dimensional inspection revealed that both curves are placed in the template shaded area.The ablation was verified by using the maestro generator 4000, and the maestro 4000 displayed a d06 error code.Electrical test was performed and the device was found out of specifications.The device has an electrical open line in the thermistor cable, high resistance on me1.X ray analysis revealed that the center support is kinked, but no evidence of the open line was noticed.The handle was opened finding no issues to the connector.The shaft was dissected finding the me1 wire damage at 60cm from the tip.This damage could cause the high resistance.The distal tip was dissected finding the thermistor wire twisted, broken and has a char on the broken area at 3.5 from the base of the tip.In addition the wire was twisted at 5cm from the base of the tip.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The most probable root cause is operational context as device performance was limited due to anatomical procedural factors.(b)(4).
|