Catalog Number 6541-4-804
Device Problems Degraded (1153); Sticking (1597); Material Integrity Problem (2978)
Patient Problems No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Event Date 11/19/2015
Event Type  malfunction  
Manufacturer Narrative
When completed, the evaluation summary will be submitted in a supplemental report.
 
Event Description
It is reported by the nurse of the hospital that after washing there is a sticky coating on the surface of the instruments.
 
Manufacturer Narrative
An event regarding santoprene handle degradation was reported.The event was confirmed.Method & results: device evaluation and results: visual analysis confirmed the reported event.The handle was returned degraded.Medical records received and evaluation: not performed because patient factors did not contribute to the reported event.Device history review: review of the device history records indicates all devices accepted into final stock met specifications.Complaint history review: there has been one other events for the lot referenced.Conclusions: a capa was initiated because a series of complaints were received for triathlon instrumentation where green santoprene over-mold handles have been reported to be degrading, becoming sticky and unstable.This event was determined to be under the scope of capa.Chemical and cytotoxicity tests showed the degraded santoprene is non-toxic.The ability to clean and sterilize the degraded instruments has not been compromised.
 
Event Description
It is reported by the nurse of the hospital that after washing there is a sticky coating on the surface of the instruments.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEADLESS PIN EXTRACTOR
Type of Device
PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer Contact
keyla navedo
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key5302781
MDR Text Key34352496
Report Number0002249697-2015-04257
Device Sequence Number1
Product Code MBH
Combination Product (y/n)N
PMA/PMN Number
K123486
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,other
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 11/19/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/16/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number6541-4-804
Device Lot NumberMC4W01
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/22/2015
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/14/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/10/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
-
-