(b)(4).Device evaluated by manufacturer: during device analysis, it was revealed that the device has broken adhesive and fluids under rings #1 and #3.In addition, the distal section is permanently curved and it had the shaft bumped close to each ring.Functional inspection revealed that the steering knob and the tension control knob functioned properly on both lock and unlock positions.Dimensional inspection both curves are not placed in the template shaded area, the device failed the dimensional test.X ray analysis revealed that the center support is kinked.The handle was opened, and a gap was found in one steering wire terminal, however, when the distal was dissected it was release and return to the normal position.The guide coild was inspected finding no issues on it.The distal section was dissected finding the center support kinked and permanent curved.The ferrule was inspected finding no issues on it.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The most probable root cause is operational context.(b)(4).
|
Reportable based on device analysis completed on 30nov2015.It was reported that the steering tip of the catheter was broken.Two 7-110-2.5-8-8 k2 blazer prime tip was selected to treat the right ventricle.The curving mechanism or the distal curve broke on both catheters while working inside the patient's body.The broken tips were successfully taken out but the procedure was completed with the second device.No patient complications reported and patient's condition is good.However, returned device analysis revealed that ring #1, #2 & #3 have broken adhesive and fluids under them.
|