Catalog Number 05.001.204 |
Device Problems
Defective Device (2588); Device Displays Incorrect Message (2591)
|
Patient Problem
No Patient Involvement (2645)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Initial reporter's phone number: (b)(4).The actual device has been returned and is currently pending evaluation.Once reliability engineering evaluates the device, a supplemental medwatch report will be sent accordingly.
|
|
Event Description
|
It was reported by (b)(6)that during service and evaluation, it was observed that the battery charger device red led was illuminating, and there was a charging bay defect.It was noted in the service order "during testing the red led lights in the cargo bay with single battery displays error".This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of this event was unknown.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
|
Manufacturer Narrative
|
The actual device was returned for evaluation.Reliability engineering evaluated the device and the reported condition was confirmed.The assignable root cause was determined to be due to wear from normal use and servicing.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|