The product was returned for investigation and the reported failure mode was confirmed.The failure mode will be monitored for future reoccurrence.A gap between the tip and insulation was noticed.The pn and lot# etchings are not present.Also the insulation material appeared to be a non stryker part.The instrument is non monopolar, therefore an insulin scan was not performed.Ifu states "before use, ensure that there are no breaks, chips, cracks, scratches, tears, or missing/loose components on the shaft insulation, handle, or housing.Do not use the instrument if any of these defects are present as this could cause unintended electrosurgical burns and life threatening complications.If damage has occurred, discontinue use and return the instrument for repair or replacement." the probable root causes is third party repair.
|