STRYKER ENDOSCOPY-SAN JOSE 5MM RATCHETED HANDLE 33CM; ENDOSCOPIC GRASPING/CUTTING INSTRUMENT, NON-POWERED Back to Search Results
Catalog Number 0250080234
Device Problem Detachment Of Device Component (1104)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/25/2015
Event Type  malfunction  
Manufacturer Narrative
The device manufacture date is not known at this time.However, should it become available it will be provided in future reports.Additional information will be provided once the investigation has been completed.
 
Event Description
It was reported that shaft insulation pulled away from tip.
 
Manufacturer Narrative
The product was returned for investigation and the reported failure mode was confirmed.The failure mode will be monitored for future reoccurrence.A gap between the tip and insulation was noticed.The pn and lot# etchings are not present.Also the insulation material appeared to be a non stryker part.The instrument is non monopolar, therefore an insulin scan was not performed.Ifu states "before use, ensure that there are no breaks, chips, cracks, scratches, tears, or missing/loose components on the shaft insulation, handle, or housing.Do not use the instrument if any of these defects are present as this could cause unintended electrosurgical burns and life threatening complications.If damage has occurred, discontinue use and return the instrument for repair or replacement." the probable root causes is third party repair.
 
Event Description
It was reported that shaft insulation pulled away from tip.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
5MM RATCHETED HANDLE 33CM
Type of Device
ENDOSCOPIC GRASPING/CUTTING INSTRUMENT, NON-POWERED
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer (Section G)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer Contact
kimberly lynch
5900 optical court
san jose, CA 95138
4087542000
MDR Report Key5312540
MDR Text Key34603072
Report Number0002936485-2015-01129
Device Sequence Number1
Product Code OCZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K973259
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 11/25/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/18/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number0250080234
Device Lot Number¿11252015¿
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/02/2015
Is the Reporter a Health Professional? No
Date Manufacturer Received11/25/2015
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-