DEPUY SYNTHES POWER TOOLS OSCILL-SAW-ATTACHM F/532.001 532.010 532; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT
|
Back to Search Results |
|
Catalog Number 532.021 |
Device Problems
Corroded (1131); Overheating of Device (1437); Difficult or Delayed Activation (2577); Naturally Worn (2988)
|
Patient Problem
No Patient Involvement (2645)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Initial reporter¿s phone number: (b)(6).This device was returned for service; however, did not meet manufacturing specifications during pre-repair assessment.Reliability engineering evaluated the device and observed that the bearing was defective, rusted worn and difficult to move.Therefore, the reported condition was confirmed.However, the assignable root cause could not be determined.If additional information should become available, a supplemental medwatch report will be sent accordingly.
|
|
Event Description
|
It was reported by (b)(6) that during service and evaluation, it was observed that the oscillating saw attachment device had rusted bearings, worn head attachment and was difficult to operate.It was noted in the service order that the device overheated.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of this event was unknown.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
|