• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY SYNTHES POWER TOOLS OSCILL-SAW-ATTACHM F/532.001 532.010 532; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY SYNTHES POWER TOOLS OSCILL-SAW-ATTACHM F/532.001 532.010 532; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT Back to Search Results
Catalog Number 532.021
Device Problems Corroded (1131); Overheating of Device (1437); Difficult or Delayed Activation (2577); Naturally Worn (2988)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
Initial reporter¿s phone number: (b)(6).This device was returned for service; however, did not meet manufacturing specifications during pre-repair assessment.Reliability engineering evaluated the device and observed that the bearing was defective, rusted worn and difficult to move.Therefore, the reported condition was confirmed.However, the assignable root cause could not be determined.If additional information should become available, a supplemental medwatch report will be sent accordingly.
 
Event Description
It was reported by (b)(6) that during service and evaluation, it was observed that the oscillating saw attachment device had rusted bearings, worn head attachment and was difficult to operate.It was noted in the service order that the device overheated.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of this event was unknown.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OSCILL-SAW-ATTACHM F/532.001 532.010 532
Type of Device
INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT
Manufacturer (Section D)
DEPUY SYNTHES POWER TOOLS
hauptstrasse 24
waldenburg CH443 7
SZ  CH4437
Manufacturer (Section G)
DEPUY SYNTHES POWER TOOLS
hauptstrasse 24
waldenburg CH443 7
SZ   CH4437
Manufacturer Contact
linda plews
1302 wrights lane east
west chester, PA 19380
6107195000
MDR Report Key5313043
MDR Text Key34694703
Report Number8030965-2015-12444
Device Sequence Number1
Product Code HWE
Combination Product (y/n)N
Reporter Country CodeSZ
PMA/PMN Number
PEXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Service and Testing Personnel
Type of Report Initial
Report Date 11/25/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/20/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number532.021
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Date Manufacturer Received12/08/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/25/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-