MAUDE Adverse Event Report: TERUMO CARDIOVASCULAR SYSTEMS CORP. TERUMO ADVANCED PERFUSION SYSTEM 1; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS-ADVANCED PERFUSION SYSTEM 1

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CARDIOVASCULAR SYSTEMS CORP. TERUMO ADVANCED PERFUSION SYSTEM 1; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS-ADVANCED PERFUSION SYSTEM 1 Back to Search Results
Model Number 816300
Device Problem Improper Device Output (2953)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 04/15/2015
Event Type  malfunction  
Manufacturer Narrative
(b)(4).This complaint is related to mdr #1828100-2015-01049.The user facility¿s biomedical engineer (biomed) stated the reported issue started in (b)(6) 2015.When we were pressing at a function, a different function was getting activated.Later the problem deteriorated gradually and it¿s screen was completely black after three months.Nothing is visible now (refer to mdr #1828100-2015-01049).
 
Event Description
It was reported that during set-up of the device for a cardiopulmonary bypass procedure, the touch screen on the central control monitor (ccm) was not working.The device was not changed out, as the customer is running the system without ccm.The surgical procedure was completed successfully.There was no delay, no blood loss, nor adverse consequences to the patient.
 
Manufacturer Narrative
The reported complaint was not verifiable.The manufacturer's subsidiary reported that their distributor replaced the central control monitor (ccm).Testing results were not available, and device was not returned to the manufacturer for laboratory analysis.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TERUMO ADVANCED PERFUSION SYSTEM 1
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS-ADVANCED PERFUSION SYSTEM 1
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORP.
6200 jackson road
ann arbor MI 48103
Manufacturer Contact
jan winder
6200 jackson road
ann arbor, MI 48103
7346634145
MDR Report Key5315448
MDR Text Key34580584
Report Number1828100-2015-01048
Device Sequence Number1
Product Code DTQ
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
K022947
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 03/29/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/21/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number816300
Device Catalogue Number816300
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/25/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/23/2008
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction Number1828100-10/14/11-020-C
Patient Sequence Number1
-
-