TERUMO CARDIOVASCULAR SYSTEMS CORP. TERUMO ADVANCED PERFUSION SYSTEM 1; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS-ADVANCED PERFUSION SYSTEM 1
|
Back to Search Results |
|
Model Number 816300 |
Device Problem
Improper Device Output (2953)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 04/15/2015 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).This complaint is related to mdr #1828100-2015-01049.The user facility¿s biomedical engineer (biomed) stated the reported issue started in (b)(6) 2015.When we were pressing at a function, a different function was getting activated.Later the problem deteriorated gradually and it¿s screen was completely black after three months.Nothing is visible now (refer to mdr #1828100-2015-01049).
|
|
Event Description
|
It was reported that during set-up of the device for a cardiopulmonary bypass procedure, the touch screen on the central control monitor (ccm) was not working.The device was not changed out, as the customer is running the system without ccm.The surgical procedure was completed successfully.There was no delay, no blood loss, nor adverse consequences to the patient.
|
|
Manufacturer Narrative
|
The reported complaint was not verifiable.The manufacturer's subsidiary reported that their distributor replaced the central control monitor (ccm).Testing results were not available, and device was not returned to the manufacturer for laboratory analysis.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
|
|