It was reported the procedure was to treat a long chronic totally occluded lesion in the mid to proximal superficial femoral artery (sfa).The supera was unpacked and prepped per instructions for use (ifu).Pre-dilatation was performed with a 6.0 x 100 mm balloon catheter.The balloon was inflated twice to 12 atmospheres for 2 minutes each time.The supera was advanced and initial deployment was appropriate; however, during deployment, the stent massively elongated into the introducer sheath.The physician could not control the delivery system during deployment to pack or compress the supera stent appropriately.The thumb slide was fully retracted to the start position and both the system and deployment levers locked prior to removal and during this step the tip separated within sheath and delivery system was stuck.The delivery system was removed with the stent 95% deployed, the separated tip in the sheath along with the guide wire as a single unit.Some resistance was observed during removal.Then, a 7f sheath was inserted to patients anatomy and another supera stent and a non-abbott stent were placed successfully to treat the lesion.There was no adverse patient effect and no clinically significant delay in the procedure.No additional information was provided.
|
(b)(4).Evaluation summary: (b)(4).The device was returned for evaluation and the reported deployment issue, elongation and difficulty removing was not able to be confirmed as the stent had already been deployed.The tip detachment was confirmed.Based on visual and functional analysis of the returned device, there is no indication of an issue/observation caused by, or related to the design, manufacture, or labeling of the device.A conclusive cause for the reported deployment difficulty and stent elongation could not be determined.The difficulty during removal and additional damage to the device is due to operational context.A review of the lot history record revealed no non-conformances.A query of the electronic complaint handling database revealed no other incidents reported from this lot.Based on the information reviewed, there is no indication the issue was caused by, or related to the design, manufacture or labeling of the device.
|